Merged Sandy Thill requested to merge st-incubator into master
All threads resolved!
@@ -15,7 +15,7 @@ redirects:
revised:
by: Sandy Thill
date: 2024-11-22
date: 2025-02-11
---
@@ -255,7 +255,7 @@ The decontamination cycle takes about 3 hours. After the decontamination itself,
![img14.png](img/img14.png)
23. Dilute the remaining H2O2 reagent in the H2O2 generator with a large volume of water and dispose of it. Rinse and wash the H2O2 generator with osmotic water.
23. Dilute the remaining H2O2 reagent in the H2O2 generator with a large volume of water and dispose of it in the sink. Rinse and wash the H2O2 generator with [osmotic water](/lab/water/#osmotic-water).
24. Place the connector cap back in the chamber. If the cap is missing, contact the Support Team.