@@ -44,7 +44,7 @@ If you are not sure a BSC is the right hood to use for your experiment, please c
<imgsrc="img/image2.jpeg"width="161"high="190">
</div>
- Be aware that **UV-lights act only on surfaces exposed to them**, the surfaces in the shadow are not disinfected and a chemical disinfection is still necessary after a UV decontamination.
- Be aware that **UV-lights act only on surfaces exposed to them**, the surfaces in the shadow are not disinfected and a chemical disinfection is still necessary after a UV decontamination. UV lights are damaging plastic material and gives a false feeling of disinfection while not being 100% efficient. Because of this, we do not recommend the usage of UV lights and advise to prefer the usage of disinfectant.
- The BSC must be **decontaminated before and after each use**. The interior surfaces should be wiped with an appropriate disinfectant that would kill any microorganisms that could be found in the cabinet. 70% ethanol is not effective against yeast and fungi, you should **use Bacillol 30 foam or tissues**.